Throughput planning in a chromatography laboratory is arithmetic, and the arithmetic usually reveals that the constraint is not the detector. A 20-minute method with 60 samples needs 20 hours of instrument time; the question is whether anyone has to be present for those 20 hours, which is a question about the autosampler.
This article sets out how to plan a batch so that it survives an unattended run. It is written by the engineering team at HENG MEI Intelligent Manufacturing.
The batch time calculation
Batch time equals sample count multiplied by cycle time, including any equilibration and wash steps. Two worked examples make the consequence obvious.
| Scenario | Samples | Cycle time | Batch time | Unattended window needed |
|---|---|---|---|---|
| Routine QC, one shift | 20 | 15 min | 5 h | Half a shift |
| Batch release, overnight | 60 | 20 min | 20 h | Overnight plus next morning |
| Stability study, weekend | 120 | 30 min | 60 h | Weekend, uninterrupted |
Where the batch time exceeds the shift, the autosampler capacity and the stability of the samples during that time become the design constraints. A tray that holds 24 vials cannot support a 60-hour run; a tray that holds 156 can, but only if the samples tolerate two and a half days in the tray.
Sample stability sets the real limit
Long unattended runs introduce a chemical problem that capacity planning ignores at its peril. Volatile solvents evaporate, light-sensitive analytes degrade, and temperature-sensitive samples drift at ambient.
- Seal vials properly. Crimp or screw caps with suitable septa; a loosely capped vial in a tray for 48 hours concentrates measurably.
- Check tray temperature. Where a cooled tray is available, use it; where it is not, plan the sequence so that sensitive samples run early.
- Protect from light. Amber vials or a covered tray for photo-labile compounds.
- Filter before loading. Particulates settle and block the needle or the column, and the failure happens at hour 30, not hour 1.

Sequence building and integration
The labour saving promised by a large tray disappears if the sequence has to be typed in by hand. Two integration questions decide this.
- Can the sequence be imported? From a spreadsheet or a LIMS export, in the sampler's accepted format. A 156-sample tray with manual entry takes an hour of clerical work and introduces transcription errors.
- Who triggers whom? Establish whether the autosampler signals the chromatograph or the reverse, and confirm the handshake with the system vendor before installation. A mismatch discovered at commissioning costs days.
Handling priority samples mid-run
Rush samples arrive during long batches. Two approaches exist: a secondary tray that accepts an insertion without disturbing the running sequence, or sequence editing that places the sample next in the queue. Both require that the software supports the operation without aborting the current run — a capability worth demonstrating before purchase, since "supports sequence editing" means different things across vendors.
Carry-over and wash
In a long sequence, carry-over accumulates silently. The wash protocol between injections should be specified for the analytes in use: volume, solvent and the number of rinse cycles. Where the method is sensitive to carry-over, a blank injection at the start and end of the tray is a cheap way to detect it — and it costs two positions in the plan.
Capacity decisions in summary
| Laboratory pattern | Reasonable capacity | Reason |
|---|---|---|
| Short methods, reloaded twice a day | 24 positions | Batch fits within a shift; reload is not a cost |
| One long method per day | 48–60 positions | Overnight run with margin |
| Weekend stability runs, two systems | 156 positions with dual inlet | Unattended for days; one sampler feeds two instruments |
| Rush samples during routine batches | Secondary tray | Insertion without aborting the sequence |
How HENG MEI supports chromatographers
HENG MEI Intelligent Manufacturing supplies the AS-24 and AS-156 liquid autosamplers, the latter with a 156-position main tray, a 15-position secondary tray and dual-inlet operation, controlled from host software. Samplers are quoted together with the system they feed, so that trigger and sequence-import requirements are confirmed before delivery.